96 Chapter: 6
Temp basal types
Based on your preference, you can select either Percent of Basal or Insulin Rate.
Insulin rate
Insulin rate is a fixed basal in units per
hour (U/H). This temp basal type is
independent of your current basal. When
you select Insulin rate (U/H) for your
temp basal type, your pump will deliver the
fixed amount you have set for the duration
as set. The amount of your temp basal
insulin rate can be set up to your maximum
basal rate setting.
If you make changes to your normal
basal rate, your U/H temp basal is not
affected and will continue to deliver as
programmed.
NOTE - Temp Basal rate cannot exceed
maximum basal rate.
.
temp basal settings
temp basal type: Insulin rate (U/H)
duration: 4:00 hours (1:05p - 5:05p)
rate: 3.50 U/H
3.50 U/H
temp basal rate
1.50 U/H
(rate 1)
2.55 U/H
(rate 2)
1.95 U/H
(rate 3)
3.50 U/H
7:00 a.m.
(07:00)
3:00 p.m.
(15:00)
12:00 a.m.
(00:00)
1:05 p.m.
(13:05)
5:05 p.m.
(17:05)