Carb ratios
Maximum ratio settings Range Warning limits
8 3 - 150 grams/u less than 5 or greater than 50 grams/u
0.1 - 5.0 u/exch less than 0.3 or greater than 3.0 u/exch
Carb units
The food entry when using Bolus Wizard feature
grams: 0 - 300 (increments: 1 gram)
exchanges 0.0 - 20 (increments: 0.5 exch)
Daily totals
Maximum records shown: 31 days of data, maximum display: 999.95 units/day. Daily delivery accuracy:
+0/-0.05 units.
Default screen
The HOME screen. For most screens, if no buttons are pressed for 30 seconds, the pump returns to this
screen.