Strona 1 - User Manual

Tetra Personal Multigas MonitorUser ManualM07237May 2010Issue 10

Strona 2 - Safety information:

Introduction TetraI. IntroductionThank you for purchasing the new Tetra Personal Multigas Monitor. Tetra is a portable multigas detector, designed to

Strona 3 - Contents

unit, allowing you to swap sensors between units or configure your unit, as appropriate, to your current needs. Plug and play will ease ma

Strona 4

Operation TetraII. Operation2.1 Switch-on sequence1. Ensure the unit is in clean air.2. Switch on Press and hold the operator button until the red

Strona 5 - Battery check

9Tetra Run mode2.2 Run modeThe Tetra unit will display up to four gas readings simul-taneously on the operator display screen. A typical display with

Strona 6 - Quickstart guide

TWA alarmTetra will display the TWA alarm when the 15 minute or 8 hour time weighted average alarm threshold is passed for toxic gases.2.4 Display opt

Strona 7 - Run mode

11Tetra Logging Peak test When performing a peak test, such as a vertical entry check, previous readings can be cleared on entry to the peak display

Strona 8 - 2. In the event of an alarm

Batteries TetraIII. Batteries 3.1 Rechargeable batteriesRecharge time for the Li-ion batteries is less then 6 hours (less, if they are not fully dis

Strona 9 - 4. Additional information

example). In Run mode, the battery icon will display six bars when it is full.4. Remove the lead from the charging socket and replace the protective

Strona 10 - I. Introduction

Alarm indications TetraIV. Alarm indicationsTetra provides two instantaneous alarm levels for each installed sensor, des-ignated level 1

Strona 11 - Software

Tetra Fixing accessoriesV. Fixing AccessoriesBelt clipTetra has a strong built-in belt clip located on the back of the unit. Lifting the small lever

Strona 12 - II. Operation

Safety information: • Read and understand all instructions in the operation section of this manual before use. • Do not substitute compon

Strona 13 - 2.3 Display symbol guide

Flow sampling TetraVI. Flow samplingAttaching the flow adaptor plateTo perform manual sampling using Tetra, a flow adaptor plate must be fitted onto

Strona 14 - Peak display

Tetra Flow samplingPumped instrumentsTetra’s inbuilt pump draws sample air in through the inlet nozzle of the flow adaptor plate and out through th

Strona 15 - 2.5 Logging

Tetra gas test accessory kitThe Gas Test Accessory is a gas testing kit designed to enable gas testing of the Tetra multigas detector using a speciall

Strona 16 - III. Batteries

sensors fitted are shown with a cross beside each. Operate the Trigger regulator and apply gas to the Tetra whilst the progress bar is counting down

Strona 17 - Tetra Batteries

6.4 Cal/TestThis can arise if having selected and confirmed calibration one or more (but not all) of the sensors are not calibrate enabled. In this ca

Strona 18 - IV. Alarm indications

VII. Maintenance and CalibrationTetra is designed to operate almost maintenance free under most conditions. However, some small items of routine maint

Strona 19 - V. Fixing Accessories

always running even in Calibration or Gas Test modes - it is essential to use the C01874 'bellows' style flow plate. Pumped units with softw

Strona 20 - VI. Flow sampling

VIII. PC interface and softwareTetra can be connected to a PC using an infrared optical link. The Tetra unit has an optical communication port:

Strona 21 - Diffusion instruments

IX. i-module replacement1. Ensure you are in a non-hazardous (safe) area. Switch off the unit2. Remove any accessories, such as the flow adaptor, if

Strona 22 - 6.2 How to perform a gas test

two quick release fixings, push the sensor mounting out of the sensor plate housing, taking care to retain any seals.Installing or replacing an i-modu

Strona 23 - Tetra Field calibration

Tetra Personal Multigas MonitorContentsUnpacking ...1Quickstart guide ...

Strona 24 - 6.4 Cal/Test

X. SpecificationDimensions 122 x 128 x 57 mm (43/4 x 5 x 21/2 inches) Weight 498 g rechargeable unit, including belt clip and 4 sensors.Housing, d

Strona 25 - Calibration method

XI. Accessories and spare partsAccessory listCrowcon part number DescriptionC01841 Aspirator Plate and GasketC01846 Aspirator assembly for pumpe

Strona 26 - Ozone detector calibrations

S011422 0-500 ppm carbon monoxide COS011425 0-20 ppm sulphur dioxide SO2S011426 0-10 ppm nitrogen dioxide NO2S011427 0-20 ppm nitrogen dioxide NO

Strona 27

XII. Troubleshooting guideSymptom/ Cause Action error messageInstrument won't switch on Flat battery. Recharge or replace battery.Pump not

Strona 28 - IX. i-module replacement

Appendix: Limitations of sensorsSensor limitationsThe sensors used in Tetra have limitations common to all such gas sensors, and users sho

Strona 30 - X. Specification

UK OfficeCrowcon Detection Instruments Ltd 2 Blacklands Way, Abingdon Business ParkAbingdon Oxfordshire OX14 1DY United KingdomTel: +44 (0)1235 55770

Strona 32

1Tetra UnpackingTetra Personal Multigas MonitorThank you for purchasing the new Tetra Personal Multigas Monitor. Tetra has redefined portable gas mon

Strona 33 - XII. Troubleshooting guide

Quickstart guide TetraQuickstart guide1. Getting startedReview your Tetra unitSwitching on your unitTetra requires little setting up, follow these si

Strona 34 - Sensor limitations

Tetra warm up sequencea) The unit will test the alarm LEDs, sounder, vibration alerts, and the operator display screen. The sounder may be silenced

Strona 35

Run mode Your unit is now ready to use. Below is a typical screen display showing the unit in normal gas monitoring Run mode. Familiarise yourself

Strona 36

1. When the gas level returns to normal, press the operator button. This will reset your Tetra unit to normal Run mode. If gas levels are still in al